Common Alarms
1
1
Move your mouse over and click on the icons-plus.png to learn more about Compat Ella®'s alarms.
  • end-of-priming-500ml-1.png

    END OF PRIMING

    Compat Ella® has automatically delivered a pre-set volume of tube feed to prime the tubing. Check the tubing to ensure that it is completely filled. If additional priming is required, the auto-prime can be repeated.

  • alarm-2.gif

    END OF THERAPY

    The programmed total volume has been completely given. End administration or start a new feeding program. NB: If "VOLUME DELIVERED THIS THERAPY" is not cleared before each newfeeding program, the end of therapy alarm may sound prematurely.

  • PumpWaiting.gif

    PUMP WAITING

    Compat Ella® is in pause mode and 2 minutes have passed without any pump activity. Resume administration or switch pump off. Extend inactivity timeout by another 2 minutes by pressing icons-start.png.

  • dooropen.gif

    Door open

    Compat Ella® door is open or the door mechanism is damaged. Close the door. If damaged, contact your local Nestlé Health Science representative or Nestlé approved service provider.

  1. no-alarms.jpg
    Acknowledge
    Alarms
  2. common-alarms.jpg
    Common
    Alarms
  3. admin-set-alarms.jpg
    Administration
    Set Alarms
  4. battery-alarms.jpg
    Battery
    Alarms
  5. other-alarms.jpg
    Other
    Alarms